CALIBRATING THE DISPENSER
CALIBRATION MODE: The SLCA-7 is calibrated at the factory and does not normally need to be re-calibrated. 
To enter the Calibration Mode, Set the RUN/NIGHT/PROGRAM Switch to the PROGRAM position and then hold 
the MENU switch down for 10 seconds.
 
LED DISPLAY 
 CALIBRATION MODE 
 DESCRIPTION
MENU
(-)
(+)
1
C
r u n
MENU
(-)
(+)
1
C 4 0
0
MENU
(-)
(+)
2
C
r u n
2
C 4
MENU
(-)
(+)
3
C
r u n
MENU
(-)
(+)
3
C
4
1
Left Pump Cal. 
 
 
 
Left Pump Cal. 
 
 
 
Right Pump Cal. 
 
 
Right Pump Cal. 
 
 
Left Coffee Water Rate 
 
 
 
Left Coffee Water Rate 
 
 
 
Set PRIME-NORMAL-RINSE switch 
to PRIME position.
Press the SMALL button, left pump 
will run for 100 counts
Enter ml collected from the 100 count 
test above.
Default = 43.0 ml
Set PRIME-NORMAL-RINSE switch 
to PRIME position.
Press the LARGE button, left pump 
will run for 100 counts
Enter ml collected from the 100 count 
test above.
Default = 43.0 ml
Set PRIME-NORMAL-RINSE switch 
to RINSE position
Press the SMALL button, left pump 
dilution water will run for 20 seconds.
Enter Liters collected from 20 sec. 
test above
Default = .414