Ethernet initial connection steps
Section 5: Startup and meter configuration
 Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
Ethernet Initial Connection Material Checklist
To complete the steps in the subsections that follow, obtain the following materials and 
information:
Daniel 3410 Series Gas Ultrasonic Flow Meter Ethernet adapter cable 
(Daniel P/N 3-3400-079)
Personal computer (PC) configured as follows:
Daniel MeterLink software installed (version 3.0 or higher)
Ethernet LAN adapter 
Configure to automatically obtain an IP address (via DHCP)
Desired meter(s) communication configuration parameters:
IP address
serial communication parameters such as baud rate and Modbus ID (if desired)
5.3.4
Ethernet initial connection steps
1.
Power up the meter.
2.
Shutdown the PC.
3.
Plug the Ethernet adapter cable Phoenix end into the meter Field Connection Board 
connector J8 and connect the RJ-45 end into the PC Ethernet connector.
4.
Enable the Ethernet LAN connector DHCP server on the CPU Module by moving the 
DHCP (switch-1) to the 
ON
 position (see direction arrow on the CPU Module label).
5.
Power up (boot) the PC and log in to the initial Windows logon prompt.
6.
Verify the Ethernet connection status by the CPU Module “LINK” LED which should be 
on solid green.
7.
Launch Daniel MeterLink.