Configuration protection
Section 5: Startup and meter configuration
 Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
Conventions used for data points reference
This section identifies the groups of data points that must be configured, along with 
interrelations and dependencies that should be considered when configuring the meter. This 
section assumes that the Mark III meter communication parameters have been configured as 
described in 
. Therefore, in the discussions that follow, the data points are grouped in 
the same way they are displayed in Daniel MeterLink screens. Also, the groups of data points are 
presented in a sequence that is recommended for configuring a new or unconfigured flow 
meter. References to specific data points, or data point labels, are printed in bold, sans serif type. 
The following data point labels are examples:
BaudPortA
EnablePressureInput
     
Also in the sections that follow, references to options for each of the data points are printed as 
they are displayed in Daniel MeterLink, along with the actual values or the data type (e.g., string, 
integer) or units of measurement (e.g., Kelvin, meters) that may be associated with the data 
point. The actual values, data type, or units of measurement are printed in parentheses. The 
following are example data point options for 
InputPressureUnit:
Gage (FALSE)
Absolute (TRUE)
In the above example, Gage and Absolute are the data point options displayed in Daniel 
MeterLink (in a dropdown list). The actual values associated with the data point are FALSE and 
TRUE, respectively, as shown in the parentheses.
5.6.2
Configuration protection
The Mark III meter provides configuration protection which is a means of protecting the 
firmware and certain configuration data points from changing. The configuration protection is 
enabled/disabled via a CPU Board switch as described in 
. Also 
In general, the data points that are protected by configuration protection fall within these 
groups:
Meter identification
Units system
Operational (i.e., transducers control, etc.)
Meter correction parameters
Alarm parameters
AGA8 configuration
Frequency outputs configuration
Digital outputs configuration
Data points labels are used extensively in Daniel MeterLink online and printed reports, 
such as Edit/Compare Configuration.