 Reference, Installation, and Operations Manual
Section 5: Startup and meter configuration
3-9000-743 Rev S
June 2013
AGA 8 Setup page
 155
h.
Use which gas composition on GC alarm: Select which gas composition the 
Daniel Mark III Gas Ultrasonic meter will use if the GC goes into alarm. 
If Fixed value is selected, the meter will start using the fixed gas composition 
stored in the meter. 
If Last good value is selected, the meter will use the last gas composition 
collected from the GC before the GC started to report alarms.
2.
Click the 
Next
 button to continue the meter configuration. Daniel MeterLink displays 
the 
Field Setup Wizard - AGA-8 Page.
5.6.14
AGA 8 Setup page
Figure 5-17  Field Setup Wizard - AGA8 Setup page 
The 
Field Setup Wizard - AGA8
 page is only displayed if both temperature and pressure are set to 
Live
 or 
Fixed
 and 
Base condition correction
 is selected in the earlier 
Field Setup Wizard - Startup
 
page (
). Configuring AGA8 parameters is necessary for obtaining the desired 
AGA8 flow-condition compressibility calculation results and for optionally calculating the sound 
velocity (using AGA10 calculations).