Справочник Пользователя для Emerson Process Management 3420
Reference, Installation, and Operations Manual
Section 5: Startup and meter configuration
3-9000-743 Rev S
June 2013
AGA 8 Setup page
155
h.
Use which gas composition on GC alarm: Select which gas composition the
Daniel Mark III Gas Ultrasonic meter will use if the GC goes into alarm.
If Fixed value is selected, the meter will start using the fixed gas composition
stored in the meter.
stored in the meter.
If Last good value is selected, the meter will use the last gas composition
collected from the GC before the GC started to report alarms.
collected from the GC before the GC started to report alarms.
2.
Click the
Next
button to continue the meter configuration. Daniel MeterLink displays
the
Field Setup Wizard - AGA-8 Page.
5.6.14
AGA 8 Setup page
Figure 5-17 Field Setup Wizard - AGA8 Setup page
The
Field Setup Wizard - AGA8
page is only displayed if both temperature and pressure are set to
Live
or
Fixed
and
Base condition correction
is selected in the earlier
Field Setup Wizard - Startup
page (
). Configuring AGA8 parameters is necessary for obtaining the desired
AGA8 flow-condition compressibility calculation results and for optionally calculating the sound
velocity (using AGA10 calculations).
velocity (using AGA10 calculations).