Batch cycle processing
 185
6.1.3
Batch cycle processing
Batch update period
Calculation updates performed by the meter, for deriving volume and velocity, are based on 
batches of data samples collected from sequences of transducer firings. The batch update 
period is dependent upon the user specified batch update period (
SpecBatchUpdtPeriod) and 
the stack size (
StackSize) as shown in 
. The Standard batch update period is the 
default.
The actual batch update period is readable via the 
BatchUpdatePeriod data point.
Emission rates
The emission rate is the period between firing two different transducers. The stacked emission 
rate is the period between consecutive firings of a single transducer when stacking is used (i.e., 
the stack size is not set to ‘None’).
The actual emission rates used (readable via the 
EmRateActual and StackEmRateActual data 
points) are functions of the desired emission rates, meter type, firing sequence, stack size, and 
filter status.
The user specifies the desired emission rates via the 
EmRateDesired and StackEmRateDesired 
data points. If a desired emission rate is set to zero, the meter uses the fastest possible emission 
rate (which can be as short as 4 milliseconds). Otherwise, the meter uses the fastest possible 
emission rate that is not less than the desired value.
When using the Daniel MeterLink 
Update Time Wizard
 to configure the meter (as discussed in 
), 
EmRateDesired and StackEmRateDesired are set to zero to achieve the most 
transducer firings per batch update.
Table 6-1  Actual meter update period
StackSize
1 (None)
2
4
8
16
SpecBatchUpdtPeriod
Rapid
(250 ms)
0.25 sec
N/A
N/A
N/A
N/A
Standard
(1000 ms)
1 sec
1 sec
1.5 sec
3 sec
5 sec