Obtaining optional keys
Section 4: Optional features
 Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
4.1.2
Obtaining optional keys
If you do not have a valid key to enter to enable a feature, you must have the CPU board serial 
number available, which displays in the 
Meter Menu|Key Manager
 dialog and the 
Meter 
Menu|Meter Information
 dialog, when you contact Daniel to obtain the key. The keys can be 
given verbally over the phone or they can be e-mailed in a Key file for easy entry.
E-mail: tech.service@emersonprocess.com
Web: 
 
(select the “Contact Us” link)
Also see, Daniel MeterLink 
Help|Technical Support
4.1.3
Optional Ethernet key
The optional Ethernet access feature allows the meter to communicate with the Daniel 
MeterLink software via the meter’s Ethernet port. This feature is enabled/disabled via the 
Eth1Key data point. 
4.1.4
Optional data log access key
The optional data log access feature allows the reading of the meter’s hourly, daily, audit, alarm, 
and system log data using Daniel MeterLink. This feature is enabled/disabled via the 
LogAccessKey data point.
4.1.5
Optional GC interface key
The optional GC interface feature allows the meter to read gas property data (composition and 
heating value) from a Daniel gas chromatograph. The meter can use the gas composition data 
to (a) calculate AGA8 compressibilities for converting volumetric flow rate and volumes to 
standard (base) condition, (b) calculate AGA8 density for calculating mass rate and mass totals, 
and (c) to optionally calculate the AGA10 sound velocity (see below). The meter can use the GC-
reported gas heating value to calculate the energy rate and totalized energy values. This feature 
is enabled/disabled via the 
GCKey data point. 
For Ethernet communication, the optional Mark III Ethernet adapter cable (Daniel P/N 
3-3400-079) is required.
The optional GC interface feature requires the use of the Option Board serial port (Port C).