Figure 3-6
HART multidrop wiring – Internal or external power
3.4.2
Frequency output wiring
Frequency output wiring depends on whether you will use internal or external power. The following 
diagrams are examples of proper wiring for these configurations:
Figure 3-7
Frequency output wiring – Internal power
HART-compatible
host or controller
HART-compatible 
transmitter
SMART FAMILY
 
transmitter
Note: For optimum HART communication, make sure the output loop 
is single-point-grounded to an instrument-grade ground.
24 VDC loop power 
supply required for 
HART 4–20 mA 
passive transmitters
600 Ohm maximum resistance
250 Ohm minimum resistance
Model 2400S
External power
Model 2400S
Internal power
Counter
000042
Note: See Figure 3-11 for output 
voltage versus load resistance.
+
Output voltage level is +24 VDC ±3%