Chapter 3
Theory of Operation
© National Instruments Corporation
3-7
Lab-NB User Manual
The recommended DAQ rates given in Table 3-2 assume that voltage levels on all the channels
included in the scan sequence are within range for the given gain and are driven by low-
impedance sources.  The signal ranges for the possible gains are shown in Table 3-3 and
Table 3-4.  Signal levels outside the ranges shown in Table 3-3 on the channels included in the
scan sequence adversely affect the input settling time.  Similarly, greater settling time may be
required for channels driven by high-impedance signal sources.
Table 3-3.  Bipolar Analog Input Signal Range Versus Gain
Gain Setting
Input Signal Range
1
-5 V to 4.99756 V
2
-2.5 V to 2.49878 V
5
-1.0 V to 0.99951 V
10
-500 mV to 499.756 mV
20
-250 mV to 249.877 mV
50
-100 mV to 99.951 mV
100
-50 mV to 49.975 mV
Table 3-4.  Unipolar Analog Input Signal Range Versus Gain
Gain Setting
Input Signal Range
1
0 V to 9.99756 V
2
0 V to 4.99878 V
5
0 V to 1.99951 V
10
0 mV to 999.756 mV
20
0 mV to 499.877 mV
50
0 mV to 199.951 mV
100
0 mV to 99.975 mV
Analog Output Circuitry
The Lab-NB provides two channels of 12-bit D/A output.  Each analog output channel can
provide unipolar or bipolar output.  Figure 3-4 shows a block diagram of the analog output
circuitry.