Head
73
 
Hints
• An automatic bone correction allows for improved 
head image quality, without any additional post-pro-
cessing.
• In order to optimize image quality versus radiation 
dose, scans are provided within a maximum scan 
field of 300 mm with respect to the iso-center. No 
recon job with a field of view exceeding those limits 
will be possible. Therefore, patient positioning has to 
be performed accurately to ensure a centered loca-
tion of the skull.
Base
Cerebrum
kV
130
130
Effective mAs
110
110
Rotation time
1.5 sec
1.5 sec
Slice collimation
1.5 mm
2.5 mm
Slice width
3.0 mm
8.0 mm
Feed/Rotation
3.0 mm
5.0 mm
Pitch Factor
1.0
1.0
Increment
3.0 mm
8.0 mm
Kernel
H31s
H31s
CTDI
Vol
25.05 mGy
25.05 mGy
Effective dose
Male: 
0.37 mSv
 
Female: 
0.38 mSv
Male: 
0.70 mSv
 
Female: 
0.77 mSv
Contrast medium IV injection
Volume
50 – 60 ml
Flow rate
2 ml/sec.
Start delay
50 – 60 sec.