Overview and Operation
English  202052A
Check the 3M
TM 
SpotOn
TM
 Temperature Monitoring System website to ensure you have the most recent version of this document. 
www.spotontemperature.com reorder #202052A
Instructions for use
1.  Ensure the SpotOn system power supply is 
connected to the back of the control unit and 
the power supply is plugged into an appropriate 
outlet. If desired, ensure the provided monitor 
cable is connected to the back of the SpotOn 
control unit and to the YSI-400 type input on the 
patient monitor. 
2.  Use an alcohol wipe to clean and disinfect the 
patient’s forehead above the orbital ridge. Allow 
the patient’s forehead to completely dry.
3.  Connect the SpotOn temperature sensor to the 
sensor cable connector. Ensure the sensor tab 
is properly connected and is fully inserted into 
the sensor cable (Figure 7). The display on the 
control unit will change to the ready screen (See 
“Ready” on page 14) once it confirms the sensor 
connection.
4.  Remove backing liners on the SpotOn temperature 
sensor to expose adhesive (Figure 8).
5.  Position the sensor cable away from the patient’s 
face and apply the SpotOn temperature sensor 
to the patient’s forehead above the orbital ridge 
(Figure 9).
Sensor Tab
Figure 7: Connect the sensor to the 
sensor cable
Figure 8: Remove the backing liners
Figure 9: Apply the SpotOn sensor to the 
patient’s forehead