16 Troubleshooting
English  202052A
Errors
Error Screen
Condition
Cause
Action
E-xx 
E-1 through E-16, E-18 
through E-99
Sensor not properly 
connected to the sensor 
cable.
Make sure the sensor is connected to the 
sensor cable. 
Sensor failure.
Replace sensor.
Cable failure.
•  Replace sensor cable.
•  Contact 3M Patient Warming technical 
service.
E-xx 
E-17
Sensor failure.
Replace sensor.
Cable failure.
•  Replace sensor cable.
•  Contact 3M Patient Warming technical 
service.
Control unit failure.
Discontinue use of the unit. Contact a 
biomedical technician.
E-xx 
E-100 and up
Control unit failure.
Discontinue use of the unit. Contact a 
biomedical technician.