Overview of running a study 
          Resetting the dosimeter 
053-644, Rev E  
Page   35 
 
eg4 & eg3 
 
35 
 
The Edge can be reset through the dosimeter keypad or via QSP-II.  The steps below outline how 
to reset your dosimeter.  
 
 
Reset 
 
1.  Ensure the meter is Off
 
2.  Turn the meter on by pressing the on/off enter         button. 
 
 
3.  When the splash screen appears, press and hold the run/stop          and on/off enter        
buttons until the splash disappears.    
 
 
 
 
 
 
 
 
 
                                         Figure 4-5:  Splash screen 
 
4.  The Setup menu appears on the Edge.  Press the up or down arrow to select the reset icon        
(symbol). 
 
 
NOTE:  Setup menu is used to reset the dosimeter, change the contrast, and return to the 
main view. 
 
 
 
 
 
 
 
 
 
                                      
  
                                Figure 4-6:  Resetting the Edge from the keypad 
5.  Press and hold the on/off enter           button.  
 
 
A 3, 2, 1 countdown will appear, and then a shaded bar will appear indicating the dosimeter is 
being reset. 
6.   Press the down arrow until the checkmark icon is selected (denoted by the arrows).  Then, 
press the on/off enter          button to return to the measurement view screen. 
 
Edge Splash screen 
Arrows and enter key 
Indicates the valid keys 
are selected 
Checkmark icon 
Is the selection to 
exit the setup menu 
Reset icon