Appendix A:  Specifications  
  
 
Approved 9-Volt Batteries 
20 
Appendix A:  Specifications  
 
Measurements  
 
Globe Temperature, Dry Bulb Temperature, Wet Bulb Temperature, % Relative 
Humidity, and WBGT Index 
 
 
Temperatures given in Celsius or Fahrenheit 
 
 
Index (displayed stay time with selected index):  ACGIH TLV, U.S. Navy PHEL 
Charts,  Flag Conditions for U.S. Navy/Marine Corp. Ashore 
 
Data Logging mode and Event logging mode 
Data logging mode records and prints all measurements at user selected interval of 1, 
2, 5, 10, 15, 30, or 60 minutes. 
 
Event logging mode (1 second record) 128K bytes of data memory. 
 
 
Languages 
English, French, Spanish, Italian, German 
 
Sensors  
Dry Bulb Sensor- 1000 Ohm Platinum RTD 
 
Accuracy and Ranges: +/-0.5 from 0°C to 120°C (+/-0.9°F from 32°F 
to 248°F) 
Waterless Wet Bulb (Humidity) Sensor 
 
Accuracy and Ranges:  Expanded measurement uncertainty of 1.1°C 
(k=2) between 0°C and 80°C (32°F and 176°F) 
Globe Sensor- 1000 Ohm Platinum RTD 
 
Accuracy and Ranges: +/-0.5 from 0°C to 120°C (+/-0.9°F from 32°F 
to 248°F) 
Relative Humidity Sensor - Integrated circuit with capacitive polymer sensor 
 
Accuracy and Ranges: +/-5% from 20 to 95% (non-condensing) 
 
Operating Temperature Range 
Sensor Assembly:  -5
C to +100C 
Electronics:  -5
C to 60C 
 
Housing 
Designed water resistant to a light rain or mist.  If rain is frequent, best practice 
would be to remote the sensor bar and keep the instrument sheltered. 
 