Well 
Type
Well Result Symbol
Result
Interpretation
Sample
Positive
The sample is presumptive positive  
for the target pathogen.
Sample
Negative
The sample is negative for the target pathogen.
Sample
Inhibited
The sample matrix was inhibitory to the assay. A 
re-test may be required.  Refer to the troubleshooting 
section and the assay kit Product Instructions for more 
information.
Sample
Inspect
The presence or absence of the target pathogen was 
indeterminate. A re-test may be required.  Refer to 
the troubleshooting section and the assay kit Product 
Instructions for more information.
Sample
Error
No bioluminescence was detected. A re-test may be 
required.  Refer to the troubleshooting section and the 
assay kit Product Instructions for more information.
Reagent 
Control
Valid
The Reagent Control was valid.
Reagent 
Control
Invalid
The Reagent Control was invalid. A re-test may be 
required.  Refer to the troubleshooting section and the 
assay kit Product Instructions for more information.
Reagent 
Control
Inspect
The Reagent Control was indeterminate. A re-test may 
be required.  Refer to the troubleshooting section and 
the assay kit Product Instructions for more information.
Reagent 
Control
Error
No bioluminescence was detected. A re-test may be 
required.  Refer to the troubleshooting section and the 
assay kit Product Instructions for more information.
Negative 
Control
Valid
The Negative Control was valid.
Negative 
Control
Invalid
The Negative Control was invalid. A re-test may be 
required.  Refer to the troubleshooting section and the 
assay kit Product Instructions for more information.
Negative 
Control
Inspect
The Negative Control was indeterminate. A re-test may 
be required.  Refer to the troubleshooting section and 
the assay kit Product Instructions for more information.
Negative 
Control
Error
No bioluminescence was detected. A re-test may be 
required.  Refer to the troubleshooting section and the 
assay kit Product Instructions for more information.
Matrix 
Control
Valid
The Matrix Control was valid.
Matrix 
Control
Inhibited
The sample matrix was inhibitory to the Matrix Control. 
A re-test may be required.  Refer to the troubleshooting 
section and the assay kit Product Instructions for more 
information.
Matrix 
Control
Error
No bioluminescence was detected. A re-test may be 
required.  Refer to the troubleshooting section and the 
assay kit Product Instructions for more information.