Troubleshooting
Troubleshooting the 3M Molecular Detection Instrument
Problem
Possible Cause
Solution
Instrument remains in 
Heat state (orange status 
light)
Core Temp shows - 
Infinity °C
Self-Check fails Core 
Heater and/or Lid Heater 
test
Pluggable heater is not installed 
correctly
Reinstall pluggable heater and power cycle the 
instrument
Puggable heater was recently 
cleaned and not completely dry
Allow heater to air dry longer
Heater pogo pins are not straight
Straighten pins
Pin heads on the inner shell are 
dirty or corroded
Clean pins
Core Temp shows “NaN”
Software and instrument are out 
of sync
Power cycle the instrument
Self-Check fails 
Photodiodes test
Instrument lid open and/or tray/
tubes inside instrument during 
self check
Remove tray/tubes, close lid, and restart 
Self-Check
Self-Check fails Power 
test
Power connections are not secure
Turn power switch to off, check connections, 
and turn switch to on.
Self-Check fails SRAM 
test
Memory corruption
Reinstall firmware. If failure persists, return 
instrument for repair.
Self-Check fails Firmware 
test
Memory corruption
Checksum mismatch
Reinstall firmware. If failure persists, return 
instrument for repair.
Self-Check fails Serial 
Number test
Invalid serial number
Reinstall firmware. If failure persists, return 
instrument for repair.
Lid does not open
Pluggable heater and/or Speed 
Loader Tray are not seated 
properly
1. Power cycle the instrument. 2. If lid not does 
open using the main lid button, use emergency 
lid release on the bottom of instrument. Remove 
the screw covering the button, and then insert 
a thin object (e.g., unfolded paperclip) into the 
hole and press the button. 3. Check that heater 
and/or tray are secure.
Solenoid malfunction
1. Power cycle the instrument. 2. If lid not does 
open using the main lid button, use emergency 
lid release on the bottom of instrument.  Remove 
the screw covering the button, and then insert 
a thin object (e.g., unfolded paperclip) into the 
hole and press the button. 3. Open lid, hold lid 
plunger down, and press lid open button. 4. If 
solenoid does not actuate (no clicking sound), 
return instrument for repair.