Philips M1351A 用户手册

下载
页码 189
Monitor Control and Display Panel
8
Chapter 1    General Information    
5. US1 Signal Quality Indicator. Indicates the quality of the signal detected by the US1 
transducer. 
6. US1 Speaker Light. On when you are hearing the US1 heartbeat. 
7. US1 Volume Keys. Sets the volume and selects the US1 heartbeat. 
8. Toco Display. Shows uterine activity. 
9. Toco Baseline Key. Zeroes the Toco display and trace to 20 units. 
10. US2 Display. Shows the FHR detected by the US2 transducer. 
11. US2 Signal Quality Indicator. Indicates the quality of the signal detected by the US2 
transducer. 
12. US2 Speaker Light. On when you are hearing the US2 heartbeat. 
13. US2 Volume Keys. Sets the volume and selects the US2 heartbeat.
M1353A Model   
1. Monitor On/Off Light. 
2. Telemetry Indicator. On when the Fetal Telemetry Receiver is connected and 
switched on. 
3. Function Key. Used to switch Twins Offset, Logic, FMP, and Fetal Alerting on and 
off. 
4. US Display. Shows the FHR detected by the US transducer. 
5. US1 Signal Quality Indicator. Indicates the quality of the signal detected by the US1 
transducer: 
6. US1 Speaker Light. On when you are hearing the US1 heartbeat. 
7. US1 Volume Keys. Sets the volume and selects the US1 heartbeat. 
8. Toco Display. Shows uterine activity. 
9. Toco Baseline Key. Zeroes the Toco display and trace to 20 units (when monitoring 
externally) or 0 units (when monitoring internally). 
10. US2/ECG Display. Shows the FHR detected by the ECG transducer. 
11. US2/ECG Signal Quality Indicator. Indicates the quality of the signal detected by the 
ECG transducer. 
12. US2/ECG Speaker Light. On when you are hearing the ECG heartbeat. 
13. US2/ECG Volume Keys. Sets the volume and selects the ECG heartbeat.