Hanna Instruments hi 93741 用户手册
19
HI 93743
Iron LR & pH
The HI 93743 meter measures the pH and the Iron (Fe) content in
water and wastewater.
water and wastewater.
SPECIFICATIONS
Range
Fe 0 to 400 µg/L
pH 5.9 to 8.5
Resolution
Fe 1 µg/L
pH 0.1
Accuracy
Fe ±10 µg/L ±8% of reading
pH ±0.1
Typical EMC
Fe ±1 µg/L
Deviation
pH ±0.1
Light Source
Light Emitting Diode @ 555 nm
Method
For Iron, Adaptation of the TPTZ Method. The
reaction between iron and the reagent causes a
blue tint in the sample.
For pH, Phenol red method. The reaction be-
tween pH and the reagent causes a red tint in
the sample.
REQUIRED REAGENTS
Code
Unit
Description
Quantity
HI 93746-0
Iron LR
TPTZ Reagent
2 packets
HI 93710-0
pH
Phenol red
5 drops
HI 93743
HI 93743
HI 93743
HI 93743
HI 93743
Iron Low Range & pH
Iron Low Range & pH
Iron Low Range & pH
Iron Low Range & pH
Iron Low Range & pH
34
FREE CHLORINE MEASUREMENT
FREE CHLORINE MEASUREMENT
FREE CHLORINE MEASUREMENT
FREE CHLORINE MEASUREMENT
FREE CHLORINE MEASUREMENT
H
I 9
3
7
0
1
• Fill the cuvet up to the mark with
10 mL of unreacted sample and
replace the cap.
replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
positioned securely into the groove.
• Press ZERO and "SIP" will ap-
pear on the display.
• Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
ready for measurement.
• Remove the cuvet.
Powder reagents procedure
• Add the content of one packet of HI 93701 DPD
Powder reagents procedure
• Add the content of one packet of HI 93701 DPD
reagent. Replace the cap and shake gently.
• Reinsert the cuvet into the instrument.
• Press READ and the display will
show "SIP" during measure-
ment.
ment.
• The instrument directly displays concentration in mg/L of free
chlorine on the Liquid Crystal Display.
Liquid reagents procedure
• Add 3 drops of HI 93701A-F DPD1 indicator and 3 drops of
• Add 3 drops of HI 93701A-F DPD1 indicator and 3 drops of
HI 93701B-F DPD1 buffer to another cuvet. Shake gently before
adding 10 mL of unreacted sample. Replace the cap and shake
gently again.
adding 10 mL of unreacted sample. Replace the cap and shake
gently again.
ZERO
Free Chlorine
10 mL
x 3
x 3