The HI 93710 meter measures both pH and chlorine (Cl
2
, free and
total) content in water and wastewater in the following ranges:
pH
5.9 to 8.5 pH units
Free Chlorine
0.00 to 2.50 mg/L (ppm)
Total Chlorine
0.00 to 3.50 mg/L (ppm).
SPECIFICATIONS
Range         pH 5.9 to 8.5
            Free Cl
2
0.00 to 2.50 mg/L
            Total Cl
2
0.00 to 3.50 mg/L
Resolution
0.1 pH/0.01 mg/L Cl
2
Accuracy
±0.1 pH
±0.03 mg/L ±3% of Cl
2
 reading
Typical EMC
±0.2 pH
Deviation
±0.02 mg/L Cl
2
Light Source
Light Emitting Diode @ 555 nm
Method
Adaptation of the EPA recommended DPD method
330.5 for chlorine analysis. The reaction with
reagents causes a pink tint in the sample.
For pH, Phenol red method. The reaction with
reagents causes a red tint in the sample.
REQUIRED REAGENTS
Code
Unit
Description
Quantity
HI  93710-0
pH
Phenol red
5 drops
HI  93701-0
Free Cl
2
DPD
1 packet
HI 93710
pH/Chlorine
46
pppppH MEASUREMENT
H MEASUREMENT
H MEASUREMENT
H MEASUREMENT
H MEASUREMENT
• Fill the cuvet up to the mark with 10 mL of
unreacted sample and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will appear on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed and
ready for measurement.
• Remove the cuvet and add 5
drops of the HI 93710 Phenol
Red Indicator. Replace the cap
and swirl the solution.
• Reinsert the cuvet into the instrument.
• Press the READ key and “SIP” will appear on the display during
measurement.
• The instrument directly displays the pH measured value on the
Liquid Crystal Display.
ZERO
pH
10 mL
x 5