Hanna Instruments hi 93732 用户手册

下载
页码 8
8
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
& SPECIFICATIONS
& SPECIFICATIONS
& SPECIFICATIONS
& SPECIFICATIONS
& SPECIFICATIONS
Range
0.0 to 10.0 mg/L (ppm)
Resolution
0.1 mg/L
Accuracy
±0.2 mg/L ±3% of reading
Typical EMC
± 0.1 mg/L
Deviation
Light Source
Light Emitting Diode @ 430 nm
Method
Adaptation of the Standard Methods for the
Examination of Water and Wastewater (18
  t h
edition), 
azide modified Winkler method. The
reaction between dissolved oxygen and the re-
agent causes a yellow tint in the sample.
Light Life
Life of the instrument
Light Detector
Silicon Photocell
Environment
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours
Auto-Shut off
After 10' of non-use
Dimensions
180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight
290 g (10 oz.)
9
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
INITIAL PREPARATION
HI 93732N is supplied complete with a 9 V
battery. Slide off the battery compartment
cover on the back of the meter and install
the battery while paying attention to its
polarity.
REQUIRED REAGENTS
Code
Description
Quantity
HI  93732A-0
Reagent A
5 drops
HI  93732B-0
Reagent B
5 drops
HI  93732C-0
Reagent C
10 drops
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
• When the LCD displays "- - -", it is ready.
• Fill one 60 mL BOD bottle completely with the unreacted
sample.
• Replace the cap and ensure that a small part of the sample spills
over. This is to make sure that no air bubbles have been trapped
inside.
• Remove the cap and add 5 drops
of HI 93732A and 5 drops of
HI 93732B. Replace the cap again
and ensure that a part of the
sample spills over.
• Swirl the solution. The sample becomes orange and a flocculant
agent will appear.
ON
OFF
9V
SLIDE
OFF