Hanna Instruments hi 93732 用户手册
8
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
& SPECIFICATIONS
& SPECIFICATIONS
& SPECIFICATIONS
& SPECIFICATIONS
& SPECIFICATIONS
Range
0.0 to 10.0 mg/L (ppm)
Resolution
0.1 mg/L
Accuracy
±0.2 mg/L ±3% of reading
Typical EMC
± 0.1 mg/L
Deviation
Light Source
Light Source
Light Emitting Diode @ 430 nm
Method
Adaptation of the Standard Methods for the
Examination of Water and Wastewater (18
Examination of Water and Wastewater (18
t h
edition),
azide modified Winkler method. The
reaction between dissolved oxygen and the re-
agent causes a yellow tint in the sample.
Light Life
Life of the instrument
Light Detector
Silicon Photocell
Environment
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours
Auto-Shut off
Auto-Shut off
After 10' of non-use
Dimensions
180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight
290 g (10 oz.)
9
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
INITIAL PREPARATION
HI 93732N is supplied complete with a 9 V
battery. Slide off the battery compartment
cover on the back of the meter and install
the battery while paying attention to its
polarity.
battery. Slide off the battery compartment
cover on the back of the meter and install
the battery while paying attention to its
polarity.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93732A-0
Reagent A
5 drops
HI 93732B-0
Reagent B
5 drops
HI 93732C-0
Reagent C
10 drops
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
• When the LCD displays "- - -", it is ready.
• Fill one 60 mL BOD bottle completely with the unreacted
sample.
• Replace the cap and ensure that a small part of the sample spills
over. This is to make sure that no air bubbles have been trapped
inside.
inside.
• Remove the cap and add 5 drops
of HI 93732A and 5 drops of
HI 93732B. Replace the cap again
and ensure that a part of the
sample spills over.
HI 93732B. Replace the cap again
and ensure that a part of the
sample spills over.
• Swirl the solution. The sample becomes orange and a flocculant
agent will appear.
ON
OFF
9V
SLIDE
OFF