Instruction Manual 
 
CON 400/410 
- 47 - 
15  ADDENDUM 3: CALCULAT ING TEMPERAT URE 
C O E F F I C IE N T S  
To determine the temperature coefficient of your sample solution use this formula:  
 
Where: 
tc = Temperature coefficient 
25 = 25 °C 
C
T1
 = Conductivity at Temp 1 
C
T2
 = Conductivity at Temp 2 
T
1
 = Temp 1 
T
2
 = Temp 2 
 
NOTE:
 A controlled temperature water bath is ideal for this procedure.  
 
1. 
Immerse the probe into a sample of your solution and adjust the temperature coefficient 
to 0% (that is, no compensation) by performing the following: 
A. 
From measurement mode, press the SETUP key to enter Setup mode.  
B. Press 
the 
MI/
S or MR/T key until the lower display reads P5.0.  
C. Press 
the 
ENTER key twice. The lower display reads tCO and the upper display 
shows the temperature coefficient value.  
D. Press 
the 
MR/
T key until the upper display shows 0.0.  
E. Press 
ENTER key to confirm the value.  
F. Press 
CA/MEAS key twice to return to measurement mode.  
2. 
Wait for 5 minutes. Note T
1
 and C
T1
 (conductivity at T
1
).  
3. 
Condition the sample solution and probe to a temperature (T
2
) that is about 5 °C to 10 °C 
different from T
1
, and note the conductivity reading C
T2
.  
 
NOTE:
 Record your results for future reference. Ideally T
1
 and T
2
 should bracket your 
measurement temperature, and should not different by more than 5 °C.  
 
4. 
Calculate the temperature coefficient of your solution according to the formula shown 
above.  
5. 
Enter the temperature coefficient you calculated into the meter.  
A. 
From measurement mode, press the SETUP key to enter Setup mode.  
B. Press 
the 
MR/
T key until the lower display reads P5.0.  
C. Press 
the 
ENTER key twice. The lower display reads tCO and the upper display 
shows the temperature coefficient value (should be 0, as per step 1 above).  
D. Press 
the 
MI/
S key until the upper display shows your calculated temperature 
coefficient.  
E. Press 
ENTER key to confirm the value.  
F. Press 
CAL/MEAS key twice to return to measurement mode.  
 
The calculated temperature coefficient will not be applied to all the meter readings.