To Output Basic Waveform 
To Output Sine Waveform 
 
Press Src1 Setup to enter the waveform setting interface. Press Wave to select 
“Sine”, at this point, you can set relative parameters of the sine signal. 
   
1.  Output   
Press Output to turn the signal output on or off. 
 
2.  To set frequency 
Press Frequency to set the frequency of the current signal and the range 
available is from 100 mHz to 25 MHz. Note that different waveform has different 
frequency, the frequency ranges of square and pulse are 100 mHz to 15 MHz; the 
frequency range of ramp is 100 mHz to 100 kHz. 
Method one: 
Rotate 
  directly to set the desired frequency value. 
Method two: 
Press 
  to input desired paramrter values and unist directly using the pop-up 
numeric keyboard. 
 
3.  To set amplitude   
Press Amplitude to set the amplitude of the current signal and the range 
available is from 20 mVpp to 5 Vpp (if the impedance is “HighZ”) or from 10 
mVpp to 2.5 Vpp (if the impedance is “50 Ω”). 
Method one: 
Rotate 
  directly to set the desired implitude value. 
Method two: 
Press 
  to input desired paramrter values and unist directly using the pop-up 
numeric keyboard. 
 
4.  To set the DC offset voltage 
Press Offset to set the DC offset of the current signal and the range available is 
from “-2.5 V + the amplitude value currently set/2” (if the impedance is “HighZ”) 
to “2.5 V + the amplitude value currently set/2” (if the impedance is “50 Ω”). 
Rotate 