Rigol DS1074B 4-channel oscilloscope, Digital Storage oscilloscope, DS1074B 用户手册

产品代码
DS1074B
下载
页码 164
RIGOL 
©  2008 RIGOL Technologies, Inc. 
User‟s Guide for DS1000B Series 
 
3-3 
Example 3: Capture a Single-Shot Signal 
To capture a single event, it needs to gather some pre-test knowledge of the signal 
in order to set up the trigger level and slope correctly. For example, if the event is 
derived from TTL logic, a trigger level of 2 volts should work on a rising edge. 
The following steps show how to use the oscilloscope to capture a single event. 
1.    Set the probe and the channel attenuations to 10X.
 
2.    Set up the trigger. 
1)  Press the MENU button in the Trigger control area to display the menu. 
2)  Press Edge to select the trigger mode. 
3)  Press Slope to select . 
4)  Press Source to select CH1. 
5)  Press Sweep to select Single. 
6)  Press Set UpCoupling to select DC. 
3.  Turn the vertical and horizontal 
  knobs to adjust the Volts/Div and 
the Time base in a proper range for the signal. 
4.  Turn the 
  knob to adjust trigger level. 
5.  Press RUN/STOP button to start capturing. When the trigger conditions are met, 
data appears on the display representing the data points that the oscilloscope 
obtained with one acquisition. 
This function helps to capture the occurrence easily, such as the noise with large 
amplitude; and set the trigger level higher a little above the normal level and press 
RUN/STOP and wait. When noise occurs, the instrument will record the waveform 
before and after the trigger. Adjust the 
  knob in the horizontal control 
area and change the level of the trigger position, will get the inverted delay trigger. 
It is useful to observe the waveform before the occurrence of the noise.