Example 9: Pass/Fail Test 
The Pass/Fail Test is one of enhanced special functions based on DS1000B series. In 
this  test  function  the  oscilloscope  compares  the  input  signal  to  the  established 
waveform  mask.  If  the  waveform  “touches”  the  mask,  a  “Fail”  signal  occurs, 
otherwise the test passes. When needed, a programmable output can be used for 
external automatic control applications. The output is built in as a standard feature, 
and is photo electricity isolated. 
 
Execute the steps as following: 
1.  Press Utility→Pass/Fail. 
2.  Press Enable Test and select ON
3.  Press Mask SettingLoad
4.  Press Load to recall the saved mask or press X Mask and Y Mask to adjust the 
horizontal limit and vertical limit with the multifunctional knob    then press 
Create Mask to create a new mask. 
5.  Press Output to select the expected outputting waveforms. 
6.  Press Operate to start the test. 
 
Figure 3-11 
Waveform display