Gw Instek GSP-830Spectrum analyzer, spectrum analyzerBandwidth (RBW) 3/30/300 kHz/4 MHz GSP-830 数据表

产品代码
GSP-830
下载
页码 181
 
 
GSP-830 User Manual 
160
 
Command Syntax 
The commands are partially compatible with IEEE488.2 (1992) and SCPI 
(Standard Commands for Programmable Instrumentation) (1994) standards. 
Commands are NON-case sensitive. 
 
Example command 
meas:freq:cen _ 9 _ khz
1
2 3 4
5
 
 
1: Command Header
 
2: Single space
 
 
3: Parameter1 
4: Single space 
 
5: Parameter2 
 
Command Header 
Several command header elements (nodes) can be 
concatenated to form a complex command. 
The above example can be separated into: 
meas: (root node) + freq: + cen: 
0/1 
0 or 1. 
Parameter example 
1to4 
Integer between 1, 2, 3, or 4. 
 0.01to5 
Decimal number between 0.01 and 5. 
 khz 
Unit (non-case sensitive) 
Message 
Terminator 
Marks the end of  a command line. Any of  the following 
is acceptable. 
 CR^END 
Line feed code (hexadecimal 0D) with 
END message 
 CR 
Line feed code 
 <dab>^END 
Last data byte with END message 
Message Separator  ; (semicolon) 
Command separator.