   AMPLITUDE 
   57
 
Set Input Impedance 
Select input impedance (50Ω/75Ω) 
 
Background 
In most cases, the default 50Ω is appropriate. Use 75Ω 
when specifically required, such as in cable TV signals. 
1.  Press the Amplitude key. 
Amplitude
 
2.  Press F6 (More). 
More
F   6
 
Panel operation 
3.  Press F1 (Input Z 50Ω/75Ω) to 
select the impedance. 
 
 
4.  When 75Ω is selected, the 
amplitude icon appears at the 
bottom of  the display. 
 
Set impedance offset (75Ω only) 
 
Background 
Impedance transformation to 75 Ω is also available 
through external devices such as impedance converter 
module (optional accessory ADP-101). In these cases an 
external loss will be induced. Impedance offset can 
compensate this effect. 
1.  Press the Amplitude key. 
Amplitude
 
2.  Press F6 (More). 
More
F   6
 
Panel operation 
3.  Make sure 75Ω is selected in F1 
(Input Z). 
 
 
4.  Press F2 (Input Z Cal). 
Input Z Cal
5.9dB
F   2
 
 
5.  Enter the offset using the 
Numerical keys and dB or Enter 
key. 
BK  SP
dB
Enter
GHz /
Sec
mSec
MHz /
kHz /
   Sec
μ
 
Range 
−10dB to +10dB, 0.1dB resolution