AUTOSET 
   61
 
RBW/VBW/Sweep 
after Autoset 
 
All the three BW related parameters, RBW, VBW, and 
Sweep, will be reset to Auto mode when using Autoset, 
regardless of  their previous settings. 
Limit vertical search range 
 
Background 
You can set the amplitude floor so that the signals lower 
than the setting will be ignored by the Autoset search. 
1.  Press the Autoset key. 
Autoset
 
Panel operation 
2.  Press F2 (Amp. Floor) to switch 
the range from Auto (whole 
range) to manual (limited range).
Amp.Floor
Auto   Man
F   2
 
 
3.  Enter the amplitude in dB using 
the Numerical keys and dB or 
Enter key. The Command 
window shows the setting. 
BK  SP
dB
Enter
GHz /
Sec
mSec
MHz /
kHz /
   Sec
μ
 
Amp. : -30dBm
 
Range dBm 
−80 to +20dBm, 0.1dB resolution 
 dBmV 
−33.01 to +66.99dBmV, 0.01dB resolution 
 dBuV 
+26.99 to +126.99dBuV, 0.01dB resolution 
 
See page50 for selecting amplitude units. 
Limit horizontal view range 
 
Background 
You can change the frequency span limit in the display to 
get a better view of the Autoset result. By default, the 
frequency span after Autoset is set at 3MHz (Auto). 
1.  Press the Autoset key. 
Autoset
 
Panel operation 
2.  Press F3 (Span) to switch the 
range between Auto (3MHz 
fixed limit) to Manual. 
Span
Auto  Man
F   3
 
 
3.  Enter the frequency using the 
Numerical and Unit keys. 
BK  SP
dB
Enter
GHz /
Sec
mSec
MHz /
kHz /
   Sec
μ
 
Span : 10MHz
 
Range 
Zero span, 2kHz to 3GHz (Manual) 3MHz fixed (Auto)