Set peak threshold 
Only the peaks below the threshold amplitude will be listed in the table. 
 
Panel operation 
1.  Press the Peak Search key. 
Peak Search
 
 
2.  Press F6 (More). 
More
F   6
 
 
3.  Press F1 (Peak Table On). 
Peak Table
On      Off
F   1
 
 
4.  Press F3 (Pk Threshold On). 
Pk Threshold
On      Off
F   3
 
 
5.  A horizontal line appears on the display. The 
GSP-830 searches and lists the peaks below this 
threshold amplitude. 
 
6.  Use the Arrow keys or the 
Scroll knob to move the 
threshold line. 
 
 
Sort peak 
The peaks are sorted in ascending frequency or descending amplitude order. 
 
Panel operation 
1.  Press the Peak Search key. 
Peak Search
 
 
2.  Press F6 (More). 
More
F   6
 
 
3.  Press F2 (Peak Sort) to switch 
between frequency and 
amplitude sorting. 
Peak Sort
Freq  Amp
F   2
 
Example: amplitude 
sort