Select VBW (Video BandWidth) 
 
Background 
VBW (Video Bandwidth) defines the smoothness of  the 
trace on the display. Combined with RBW, VBW defines 
the ability to sort out target signal from surrounding 
noise or adjacent peaks. 
See page99 for suitable VBW selection. 
Panel operation 
1.  Press the BW key. 
BW
 
 
2.  Press F2 (VBW) to select Auto 
or Manual. 
VBW
Auto  Man
F   2
 
 
3.  When Man (manual) is selected, 
use the Arrow keys or Scroll 
knob to select the VBW. The 
selected VBW appears in the 
command window. 
 
 
VBW: 1kHz
 
Mode Auto 
VBW is automatically selected. See 
page99 for setting. 
 Manual 
 
VBW is manually selected. The VBW 
icon appears at the bottom of  the 
display. 
Range 
10Hz to 1MHz in 1-3 steps 
In Auto mode, VBW is automatically selected in 
combination with RBW. See page99 for the settings. Use 
the Auto mode settings for reference when manually 
selecting the VBW. 
Note 
The GSP-830 automatically changes the VBW 
according to RBW selection.