5.6 
Adjustment 
As the acceleration value due to gravity is not the same at every location on earth, 
each balance must be coordinated –  in compliance with the underlying physical 
weighing principle - to the existing acceleration due to gravity at its place of location ( 
only if the balance has not already been adjusted to the location in the factory).  This 
adjustment process must be carried out during the initial start-up, after change in lo-
cation and variation of surrounding temperature. It is also recommendable to adjust 
the balance periodically during weighing operation in order to obtain exact measured 
values. 
 
5.6.1  Adjusting 
Using a precision weight, the accuracy of the balance can be checked at any time 
and adjusted. 
Adjustment procedure: 
 
Check that the surrounding conditions are stable.  
A short warm-up time of about 3 minutes is recommended for stabilisation.  
 
• Use 
the   key to switch the balance on.  
• Press 
the   key and hold depressed for approx. 10 seconds until “CAL“ ap-
pears on the display. 
• Release 
the   key; the size of the calibration weight is shown on the display 
(see chapter 1, Technical Data). 
•  Now place the calibration weight in the centre of the weighing plate.   
•  “F“ will appear for a brief period of time before the balance switches off.   
•  Remove the calibration weight. Calibration is completed.   
 
In case of an adjusting error or a wrong adjusting weight “E“ will appear. Repeat the 
adjusting procedure. 
Keep the adjusting weight near the balance. Daily verification of the balance accu-
racy is recommended for quality assured applications. 
 
EMB-BA-e-1132 
10