6    Care, Cleaning and Maintenance 
   
The Preamplifier needs no care and maintenance. 
   Cleaning 
   
Do not use solvent to clean the housing. 
7    Troubleshooting 
   
 
Type of Malfunction 
Possible Causes 
Corrections 
No LED lights. 
– No 24 V supply. 
 
– Pressure sensor is defect. 
– Sensor cord is defect. 
– LED is defect. 
• Check connection cord, if 
necessary change it. 
• Change pressure sensor. 
• Change sensor cord. 
• Loctite service. 
Amplification is not adjustable 
It lights only the same LED´s. 
– Potentiometer R10 is defect. 
• Loctite service 
There is always an error message 
at the controller, but the LED´s 
light correct. 
– Preamplifier cord is defect. 
– Preamplifier is defect. 
• Change Preamplifier cord 
• Loctite service 
There is always an error message 
at the controller and no LED lights. 
– Amplification factor not correctly
adjusted. 
– Preamplifier cord is defect. 
– Pressure sensor is defect. 
– Sensor cord is defect. 
– Preamplifier is defect. 
• Adjust the amplification factor. 
 
• Change preamplifier cord. 
• Change pressure sensor. 
• Change sensor cord. 
• Loctite service.