Controller Configuration
The controller has a number of configuration and operating options and calibra-
tion parameters which are programmable via the front panel. These are ac-
cessed from the secondary menu after the proportional band function by
pressing “SET”. Pressing “SET” again enters the first of three sets of configu-
ration parameters: operating parameters, serial interface parameters, and cali-
bration parameters. The menus are selected using the “UP” and “DOWN” keys
and then pressing “SET”. (See Figure on page 36.)
9.9
Operating Parameters
The operating parameters menu is indicated by,
PAr
Operating parameters menu
The operating parameters menu contains the High Limit, Stir Speed, and Cool-
ing parameters.
9.9.1
High Limit
The High Limit Parameter adjusts the upper set-point temperature. The factory
default and maximum temperature are set to 100°C. For safety, a user can ad-
just the High Limit down so the maximum temperature set-point is restricted.
HL
High Limit parameter
Press “SET” to enable adjustment of HL.
HL
Flashes “HL” and then displays the setting
H=100
Current HL setting
Adjust the HL parameter using “UP” or “DOWN”.
H=90
New HL setting
Press “SET” to accept the new temperature limit.
9.9.2
Low Limit
The Low Limit (LL) Parameter adjusts the lower set-point temperature limit.
The factory default and minimum temperature are set to –90°C. For safety, a
user can adjust the Low Limit up so the minimum temperature set-point is
restricted.