Installation
5.3.2 Head Ambient Temp. / Alarm Output
This output can be configured either as output for the head ambient
temperature (default configuration) or as an alarm output.
temperature (default configuration) or as an alarm output.
Power +
Power –
Head Ambient Temp.
Ground
Figure 10: Wiring the Output for Head Ambient Temperature
The output range for the head ambient temperature is 0 to 500°C
(32 to 932°F) with 10 mV /°C.
In case of an alarm the output switches between 0 V and 5 V. The
alarm output is controlled by the target temperature or the sensing
head temperature.
(32 to 932°F) with 10 mV /°C.
In case of an alarm the output switches between 0 V and 5 V. The
alarm output is controlled by the target temperature or the sensing
head temperature.
Power -
Power +
Figure 11: Wiring of the Alarm Output