Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
8-14 
8.8. 
  Prom Replacement Procedure 
Preparation: If any setup changes such as RANGE, AUTOCAL, ON/OFF etc. have been made, 
record the changes because all settings should be checked after the PROM is changed. See  
Figure 9-2 for location of prom on CPU card. 
1.  Turn the machine off. 
NOTE 
The PROM is static sensitive, be sure that you are properly 
grounded before handling any of the electronics. 
 
2.  Remove the hold down screw that holds in the V/F-CPU assembly to the motherboard. 
Disconnect the J9 power connector from the motherboard. Gently lift the assembly far 
enough out of the instrument to remove the connector to the display and the RS-232 
connector. 
3.  Locate the CPU board, which is attached to the larger V/F board. 
4.  Remove the CPU-V/F board assy, laying it down on an insulating surface such that the board 
edge pins on the PCB are on the left. The PROM chip will be at the top center. See  
Figure 9-2 for prom location. Gently pry the chip from its socket and replace it with the new 
chip. Install the chip in the left end of the socket with the notch facing to the right. Make sure 
that all of the legs insert into the socket correctly. 
5.  Re-attach the CPU board to the V/F board, re-connect the RS-232 and front panel display 
cables, and then re-attach the assembly to the motherboard. 
6.  Turn the M200AH ON and observe the front panel display. As the machine goes through the 
setup the version number will be displayed on the front panel. It should read the same as the 
version number printed on the prom. 
7.  Check and re-enter if necessary any non-default settings such as RANGE or AUTOCAL. Re-
enter the SPAN value. Check all settings to make sure that expected setup parameters are 
present. 
8.  Re-calibrate the machine so that the default slope and intercept are overwritten with the 
correct values.