Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
9-12 
9.1.3.1.   Signal I/O Diagnostics 
Table 9-4:  Diagnostic Mode - Signal I/O 
No. 
Signal 
Control 
Description 
DSP_BROWNOUT 
NO 
Display brownout is used to keep the display from 
getting corrupted during low line voltage conditions. 
Circuitry on the Status/Temp board senses low line 
voltage and sets this bit. The CPU reads this and 
generates the BROWNOUT_RST signal described 
below. 
EXT_ZERO_CAL 
NO 
Shows state of status input bit to cause the M200AH 
to enter Zero Calibration mode. Use to check external 
contact closure circuitry. 
EXT_SPAN_CAL 
NO 
Shows state of status input bit to cause the M200AH 
to enter the Span Calibration mode. Use to check 
external contact closure circuitry. 
EXT_LOW_SPAN 
NO 
Shows state of status input bit to switch the low span 
concentration valve in the zero/span valve option. 
REMOTE_RANGE_HI 
NO 
Shows state of status input bit to switch the remote 
range control from low to high range. 
SPAN_VALVE 
YES 
Switches the Span valve. Use this bit to test the valve 
function. 
CAL_VALVE 
YES 
Switches the Cal valve. Use this bit to test the valve 
function. 
NOX_VALVE 
YES 
Switches the NO/NO
x
 valve. Use this bit to test the 
valve function. 
RCELL_HEATER 
YES 
Shows the status of the reaction cell heater. This has 
the same function as the LED in the PSM. 
BLOCK_HEATER 
YES 
Shows the status of the sample flow orifice block 
heater. This has the same function as the LED in the 
PSM. 
10 
ELEC_TEST 
YES 
Turns on electric test bit in preamp. Should be used 
for troubleshooting Preamp logic lines. We 
recommend you use the ET button in the DIAG menu 
to operate electric test.