9.3.3.1.   ADC/DAC Calibration Procedure 
Due to the stability of modern electronics, this procedure should only need to be performed if a 
major sub-assembly is exchanged or when the display voltage does not match the input voltage 
or current to the V/F card. After completion, a Factory Calibration Procedure should be 
performed, see Section 9.1.6. 
Before the actual calibration is performed, switches on the V/F card must be correctly set and 
jumpers set on the motherboard. Jumper and switch setting changes must be performed with 
the instrument power OFF.  
Motherboard Jumpers 
The motherboard contains 4 pairs of jumpers JP1 - JP8, one pair for each analog output channel. 
Each channel can be configured for either voltage or current output. Use Table 9-7 to configure 
the jumpers. 
Table 9-7:  Motherboard Jumper Settings 
Analog Output 
Terminal Pair 
Rear panel 
Jumper Pair 
Jumper Setting for 
Voltage Mode 
Jumper Setting for 
Current Mode 
DAC 0 - NO
X
 
3 - 4 
JP3 - JP4 
B - C 
A - B 
DAC 1 - NO 
5 - 6 
JP1 - JP2 
B - C 
A - B 
DAC 2 - NO2 
(units w/o O
2
1 - 2 
JP5 - JP6 
B - C 
A - B 
DAC 2 - O
2
 
(units with O
2
1 - 2 
JP5 - JP6 
B - C 
A - B 
DAC 3 - TEST 
7 - 8 
JP7 - JP8 
B - C 
A - B 
 
V/F Board Switch Settings 
There are 2 different types of current outputs, Non-Isolated (std equipment) and Isolated. Each 
requires a different switch setting shown below. If you are operating the instrument in voltage 
output mode, the switches should be set to the desired voltage range. 
Table 9-8:  V/F Board Switch Settings 
DAC # 
Sw 1 
Sw 2 
Sw 3 
Sw 4 
Sw 5 
Sw 6 
Sw 7 
ON* 
OFF* 
    
OFF* 
ON* 
OFF* 
    
OFF* 
ON* 
OFF* 
    
OFF* 
ON* 
OFF* 
    
OFF* 
*Required settings