Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
9-45 
9.3.4.4.   4-20 mA Current Output 
There are two types of current loop outputs available for the M200AH, non-isolated (standard 
equipment) and isolated (optional). These outputs may be used in place of the analog voltage 
signals that appear at the rear panel connector. See Section 9.3.3.1 for setup and calibration 
procedure.  
Non-isolated 4-20 mA or 0-20 mA Output 
Non-isolated 4-20 mA is recommended for applications requiring a current loop and the current 
loop receiver has an isolated input stage. It can also be used when there a potential difference 
does not exist between the ground connections of the two instruments.  
Recommended load (R
l
) = 100
 - 600 (Max. 600) 
V/F analog output voltage setting = +10 V Full Scale 
Default Setting = 4-20 mA 
To convert from 4-20 mA to 0-20 mA output, refer to Status/Temp Schematic 01087 sheet 2, and 
the procedure in Section 9.3.3.1. 
Isolated 4-20 mA or 0-20 mA Option 
This option is recommended where maximum isolation from ground loops is required. 
Specifications include up to 1500 V rms common mode voltage isolation (transformer coupled 
input to output galvanic isolation) and 240 Vrms normal mode voltage output protection.  
Recommended load (R
l
) = 200
 - 1000 (Max. 1200) 
V/F analog output voltage setting = +5V Full Scale 
Default Setting = 4-20 mA 
To convert from 4-20 mA to 0-20 mA output, refer to Status/Temp Schematic in the Appendix, 
and the procedure in Section 9.3.3.1. 
9.3.5.   Power Supply Module 
The Power Supply Module consists of several subassemblies described in Table 9-10.