9.3.7.
Flow/Pressure Sensor
The flow/pressure sensor board consists of 2 pressure sensors and a flow sensor. See Figure 9-8
for a diagram of this board. From these three sensors four values are computed and displayed on
the front panel TEST functions. They are:
for a diagram of this board. From these three sensors four values are computed and displayed on
the front panel TEST functions. They are:
1. Ozone Flow - measured directly - S3, R3
2. Reaction Cell Pressure - measured directly - S2, R2
3. Sample Gas Pressure - measured directly S1, R1
4. Sample Flow - computed from sample pressure and reaction cell pressure S1, S2
The above pressures and flows are filtered to produce the front panel readings. There is a small
delay after adjustment for a steady reading when observing the TEST functions.
delay after adjustment for a steady reading when observing the TEST functions.
The flow/pressure sensor board consists of 2 pressure sensors and a flow sensor. See Figure 9-8
for a diagram of this board. From these three sensors four values are computed and displayed on
the front panel TEST functions. They are:
for a diagram of this board. From these three sensors four values are computed and displayed on
the front panel TEST functions. They are:
1. Ozone Flow - measured directly - S3, R3
2. Reaction Cell Pressure - measured directly - S2, R2
3. Sample Gas Pressure - measured directly S1, R1
4. Sample Flow - computed from sample pressure and reaction cell pressure S1, S2
The above pressures and flows are filtered to produce the front panel readings. There is a small
delay after adjustment for a steady reading when observing the TEST functions.
delay after adjustment for a steady reading when observing the TEST functions.
To adjust the OZONE flow:
1. Go to DIAG mode by pressing SETUP-MORE-DIAG, then select SIGNAL I/O, and press
ENTR. Select OZONE_FLOW by using NEXT-PREV keys.
2. Adjust R3 so that OZONE_FLOW reads 2000 mV. This is the coarse adjustment.
3. Press EXIT to return to the SAMPLE mode. Select the OZONE FL - TEST function. Plumb
a calibrated flow measuring device upstream of the ozone generator (see Figure 9-7). Use the
external device reading to make small adjustments in R3 to dial in correct flow readout on
the front panel.
external device reading to make small adjustments in R3 to dial in correct flow readout on
the front panel.