Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
9-59 
9.3.8.4.   PMT Cooler 
The PMT cooler uses a Peltier cooler supplied with DC current from the switching power supply 
in the Power Supply Module. An overall view is shown in Figure 9-11. The temperature is 
controlled by a proportional temperature controller located on the Preamp board. The input 
voltage from the supply is 15 VDC. Typical current through the cooler at room temp is about 1.5 
to 2 A. At normal room temperatures the 7
C set point should be maintained within 0.1C. The 
current and voltage can be measured at test points on the fan shroud at one end of the NO
X
 
sensor module. See the Schematic in the Appendix.