9.3.8.4. PMT Cooler
The PMT cooler uses a Peltier cooler supplied with DC current from the switching power supply
in the Power Supply Module. An overall view is shown in Figure 9-11. The temperature is
controlled by a proportional temperature controller located on the Preamp board. The input
voltage from the supply is 15 VDC. Typical current through the cooler at room temp is about 1.5
to 2 A. At normal room temperatures the 7
in the Power Supply Module. An overall view is shown in Figure 9-11. The temperature is
controlled by a proportional temperature controller located on the Preamp board. The input
voltage from the supply is 15 VDC. Typical current through the cooler at room temp is about 1.5
to 2 A. At normal room temperatures the 7
C set point should be maintained within 0.1C. The
current and voltage can be measured at test points on the fan shroud at one end of the NO
X
sensor module. See the Schematic in the Appendix.