Teledyne 200AH 用户手册
Teledyne API Model 200AH NO
X
Analyzer Instruction Manual, 06492, Rev. G2
3-1
3.
SPECIFICATIONS, WARRANTY
3.1.
Specifications
NOX Specifications
O
2
Option
Operating Modes
NO/NO
x
switching mode, NO only
mode, NO
x
only mode.
Paramagnetic Sensor
Ranges
In 1 ppm increments from 5 ppm to
5000 ppm; Single range, independent
ranges or autoranging
5000 ppm; Single range, independent
ranges or autoranging
In 1 % increments from 5 % to
100%, Independent of NO
100%, Independent of NO
X
range
Noise at zero
<0.020 ppm RMS
0.05% O
2
Noise at span
<0.5% of reading RMS above 20 ppm
<0.1% O
2
Detection Limit (Note 1)
0.040 ppm
0.1% O
2
Zero Drift (Note 2)
<0.5% full scale/24 hours
0.1% O
2
Zero Drift (Note 2)
<1.0% full scale/7 days
0.1% O
2
Span Drift (Note 2)
<1% FS/24 hours
0.1% O
2
Lag Time
- Switching Mode
- 20 sec (Note 3)
- NO
x
mode
- 4 sec (Note 3)
< 4sec
Response Time
- Switching Mode
- 95% in < 40 sec (Note 3)
- NO
x
mode
- 95% in < 10 sec (Note 3)
95% in <10 sec
Sample Flow Rate (Analyzer)
370 cc/min
10% (Including bypass)
Sample Flow Rate (Sensors)
40 cc/min
10%
80 cc/min
10%
Linearity
1% of full scale
0.1% O
2
Precision
0.5% of reading
0.1% O
2
Temperature Range
5-40
C
Humidity
0-95% RH non-condensing
Temp Coefficient
< 0.1% per
C
< 0.1% per
C
Voltage Coefficient
< 0.1% per V
< 0.1% per V
Dimensions HxWxD
7"x17"x23.6" (18 cm x 43 cm x 61 cm)
Weight, Analyzer
45 lbs (22 kg)
Weight, Pump Pack
16 lbs (7 kg)
Power, Analyzer
100 V~ 50/60 Hz, 120 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 200 watts
Power, Analyzer
4
230 V~ 50 Hz, 2.5A
Power, Ext Pump
110 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 295 watts
Power, Ext Pump
4
230 V~ 50 Hz, 2.5A
Environmental
Installation Category (Over-voltage Category) II
Pollution Degree 2
Analog Resolution
1 part in 2048 of selected voltage or
current range
current range
1 part in 2048 of selected voltage or
current range
current range
Recorder Output
0-100mV, 0-1, 5, 10v, bipolar
0-100mV, 0-1, 5, 10v, bipolar
Current Loop Option
4-20ma isolated
4-20ma isolated
Status
12 Status Outputs from opto-isolator
Measurement Units
ppm, mg/m
3
%
1.
Defined as twice the zero noise level.
2.
At constant temperature and voltage.
3.
Lag & response times longer for external converter option.
4.
Electrical ratings for CE Mark compliance.