Teledyne 200AH 用户手册

下载
页码 196
Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
3-1 
3. 
  SPECIFICATIONS, WARRANTY 
3.1. 
  Specifications 
 
NOX  Specifications 
O
2
 Option 
Operating Modes 
NO/NO
x
 switching mode, NO only 
mode, NO
x
 only mode. 
Paramagnetic Sensor 
Ranges 
In 1 ppm increments from 5 ppm to 
5000 ppm; Single range, independent 
ranges or autoranging 
In 1 % increments from 5 % to 
100%,  Independent of NO
X
 range 
Noise at zero 
<0.020 ppm RMS 
0.05% O
2
 
Noise at span 
<0.5% of reading RMS above 20 ppm 
<0.1% O
2
 
Detection Limit (Note 1) 
0.040 ppm  
0.1% O
2
 
Zero Drift (Note 2) 
<0.5% full scale/24 hours 
0.1% O
2
 
Zero Drift (Note 2) 
<1.0% full scale/7 days 
0.1% O
2
 
Span Drift (Note 2) 
<1% FS/24 hours 
0.1% O
2
 
Lag Time 
 
- Switching Mode 
- 20 sec (Note 3) 
- NO
x
 mode 
- 4 sec (Note 3) 
< 4sec 
Response Time 
 
- Switching Mode 
- 95% in < 40 sec (Note 3) 
- NO
x
 mode 
- 95% in < 10 sec (Note 3) 
95% in <10 sec 
Sample Flow Rate (Analyzer) 
370 cc/min 
10% (Including bypass) 
 
Sample Flow Rate (Sensors) 
40 cc/min 
10%  
80 cc/min 
10%  
Linearity 
1% of full scale 
0.1% O
2
 
Precision 
0.5% of reading 
0.1% O
2
 
Temperature Range 
5-40
C 
Humidity 
0-95% RH non-condensing 
Temp Coefficient 
< 0.1% per 
C 
< 0.1% per 
C 
Voltage Coefficient 
< 0.1% per V 
< 0.1% per V 
Dimensions HxWxD 
7"x17"x23.6" (18 cm x 43 cm x 61 cm) 
Weight, Analyzer 
45 lbs (22 kg) 
Weight, Pump Pack 
16 lbs (7 kg) 
Power, Analyzer 
100 V~ 50/60 Hz, 120 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 200 watts 
Power, Analyzer
4
 
230 V~ 50 Hz, 2.5A 
Power, Ext Pump 
110 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 295 watts 
Power, Ext Pump
4
 
230 V~ 50 Hz, 2.5A 
Environmental 
Installation Category (Over-voltage Category) II 
 
Pollution Degree 2 
Analog Resolution 
1 part in 2048 of selected voltage or 
current range 
1 part in 2048 of selected voltage or 
current range 
Recorder Output 
0-100mV, 0-1, 5, 10v, bipolar 
0-100mV, 0-1, 5, 10v, bipolar 
Current Loop Option 
4-20ma isolated 
4-20ma isolated 
Status 
12 Status Outputs from opto-isolator 
Measurement Units 
ppm, mg/m
3
 % 
1. 
Defined as twice the zero noise level. 
2. 
At constant temperature and voltage. 
3. 
Lag & response times longer for external converter option. 
4. 
Electrical ratings for CE Mark compliance.