Teledyne API Model 200AH NO
X
Analyzer Instruction Manual, 06492, Rev. G2
5-6
5.1.1. Sample Menu
Table 5-1: M200AH Sample Menu Structure
Level 1
Level 2
Level 3
Level 4
Level 5
Description
Reference
Section
Section
TEST
TST>
TST>
Test
functions
CAL
Zero/Span calibration w/ gas through sample
port
port
CALZ
Zero calibration w/ zero gas from zero valve
option
option
CALS
Span calibration w/ span gas from low or high
concentration span valve option
concentration span valve option
LOW/
HIGH
Only present if AutoRange is selected. Used
to select which range instrument will be
calibrated in. Selects which slopes, offsets,
expected span values, etc. will be updated and
displayed.
to select which range instrument will be
calibrated in. Selects which slopes, offsets,
expected span values, etc. will be updated and
displayed.
ZERO
Press ZERO then ENTR will zero analyzer
SPAN
Press SPAN then ENTR will span analyzer
CONC
Expected
NO/NO
x
span concentrations and
Conv efficiency setup
NOX
CONC
Enter expected NO
x
span concentration
NO
CONC
CONC
Enter expected NO span concentration
CONV
Sub-menu for converter efficiency setup and
verification
verification
NO2
Expected
NO
2
concentration for converter
efficiency calculation
CAL
Automatic
converter
efficiency
calibration
and entry
SET
Set
the
converter
efficiency
manually 5.2.2.7,
SETUP
The SETUP Menu - See next table.