  Sample Mode 
5.2.1.   Test Functions 
NOTE 
In any of the following TEST functions, if a value of  
XXXX is displayed, that indicates an off scale and  
therefore meaningless reading. 
 
TEST functions allow the operator to examine several important operating parameters of the 
M200AH to verify correct Analyzer operation. If faults are discovered, refer to Troubleshooting 
Section 9. 
Range (RANGE) 
This is the Range of the instrument. In single range mode there is one range for all 3 rear panel 
analog outputs. 
The AutoRange option allows 2 different ranges for each channel, and will automatically switch 
to the other range dynamically as concentration values require. By selecting Remote AutoRange, 
the range change can be controlled remotely. The TEST values will show the range the 
instrument is currently operating in, and will dynamically display range change information 
when necessary. 
Independent range option allows different ranges for each output. When enabled, there will be 
three range values displayed, NO, NO
X
 and NO
2
NOTE 
Each of the range modes Single range, Auto range, and  
Independent ranges are mutually exclusive. 
 
Oxygen Sensor Range (O
2
 RANGE) (Option) 
For units equipped with the O
Sensor Option, this is the Full Scale Range of the rear panel O
analog output.