5.2.2.   CAL, CALS, CALZ, Calibration Functions 
5.2.2.1.   CAL, CALS, CALZ 
The CAL, CALS, and CALZ keys control the calibration functions of the analyzer. In the CAL 
mode the analyzer can be calibrated with zero/span gas coming in through the sample input port 
on the rear panel. If the analyzer is equipped with the Zero/Span valve option, the CALZ and 
CALS buttons operate the Zero/Span valves. The setup of this option is covered in Section 6.2, 
and operation is explained in Section 7. 
5.2.2.2.  Gas To Cal 
For units equipped with the Oxygen Sensor Option, depressing the CAL key will cause the 
instrument to prompt for which gas sensor to calibrate; NO
X
 or O
2
.  After the desired gas is 
selected, the instrument will display ZERO, SPAN, and CONC keys as described below. 
5.2.2.3.   Zero 
Pressing the ZERO key along with ENTR will cause the instrument to adjust the OFFSET value 
of the internal formulas so that the instrument reads zero. The M200AH allows zero adjustment 
over a limited range of signal levels, therefore the signal does not have to be exactly zero for the 
instrument to do a zero cal. The instrument will not, however, allow a zero cal on any signal 
level, therefore it is not possible to zero the instrument with high concentrations of span gas in 
the reaction cell. If the ZERO key does not come on as expected, check Section 9.2.9. 
5.2.2.4.   Span 
Pressing the SPAN key along with ENTR will cause the instrument to adjust the SLOPE value of 
the internal formula so the instrument displays the span value.  The expected NO
X
, NO span 
Like the Zero calibration, the Span cal cannot be done with any concentration of span gas. If 
signal level is outside certain limits the SPAN key will not be illuminated. If you encounter this 
condition see Section 9.2.8. It is also possible at low levels of span concentration that BOTH the 
ZERO and SPAN keys might be on, thus allowing you to either zero or span the instrument. In 
this case care must be taken to perform the correct operation or the analyzer can become miss-
calibrated.