Table 5-11:  RS-232 Terminal Mode Editing Keys 
Key 
Function 
CR (carriage return) 
Execute command 
BS (backspace) 
Backspace one character to the left 
ESC (escape) 
Erase entire line 
Control-R (ASCII 18 decimal) 
Recall previous command 
Control-E (ASCII 5 decimal) 
Recall and execute previous command 
 
NOTE 
To open the help screen, Type "?" and press the Enter key. 
 
Commands are not case-sensitive; you should separate all command elements (i.e. keywords, 
data values, etc.) by spaces. 
Words such as T, SET, LIST, etc. are called keywords and are shown on the help screen in 
uppercase, but they are not case-sensitive. You must type the entire keyword; abbreviations are 
not accepted. 
5.7.2.   Command Summary 
The information contained in the rest of this section covers all of the normal commands that are 
required to operate the instrument from a remote terminal. If you are going to be writing 
computer programs to communicate with the M200AH (i.e. operating the port in COMPUTER 
MODE) we suggest that you order a supplementary manual "The RS-232 Interface", Teledyne 
API part number 01350. This manual shows additional features of the port designed to support a 