Philips MP60/70/80/90 Benutzerhandbuch

Seite von 414
EEG Reports
18 Monitoring EEG
213
Changing Filter Frequencies
The low and high pass filters screen out undesirable interference from the raw EEG wave display. The 
current EEG filter frequency settings are shown in the header of the CSA. Changing filter settings 
affects the EEG wave and all the EEG numerics. Whenever the filter setting is changed, a filter change 
marker appears next to the spectral lines. 
To change the filter settings:
1
In the 
Setup EEG
 menu, select 
Low Filter
 or 
High Filter
 to call up a list of available 
frequencies.
2
Select the required frequency from this list. 
Changing the Speed of the EEG Wave
The EEG measurement has its own speed control and is not affected by the wave speed settings of the 
other measurements.
In the 
Setup EEG
 menu, select 
Wave Speed
. Choose the required speed from the pop-up list. 
This defines the speed at which the wave is drawn across the screen in millimeters per second (mm/
s).
EEG Reports
The content of EEG Reports is always the same and does not need to be configured. 
To print an EEG Report, in the 
Setup EEG
 menu, select 
Print Report
.
Alternatively, you can select the CSA and use the 
Print Report
 pop-up key to start the report.
To modify the buffer and trendline settings on the CSA Report, in the 
Reports
 menu, select 
CSA On EEG Rep.
. If you do not change these settings, the monitor will use the default 
settings SEF Trendlines: On, Buffer: C.
CSA on Report
This menu entry lets you
Buffer
choose one of the three pre-configured buffer times
Trend SEF/MDF/PPF
switch the trendline of the specific numeric on or off