Using VueLink Screens
27 VueLink Modules
283
Using VueLink Screens
Your monitor may be 
configured to show 
VueLink device data 
permanently on the 
Screens. 
Select the device data 
window to display the 
VueLink pop-up keys that 
let you access the setup 
menu and carry out 
VueLink tasks.
Switching VueLink On and Off
To switch VueLink measurements on and off, in the 
Setup VueLink
 menu, select 
Device Interface
 to toggle between the settings 
On
 and 
Off
Alarms/INOPs From External Devices
The VueLink module itself generates INOPs, but does not generate alarms. If the external device’s 
alarms are on, the module transmits these to the monitor. A message in the VueLink info window tells 
you either 
Device Alarms Ignored 
or
 Device Alarms Accepted 
or 
No Alarms 
Available.
 External device alarms status symbols precede some, but not all, measurement labels.
Alarms from external devices are:
• all transmitted to the monitor. For all numerics configured in the 
Setup VueLink
 menu, an 
alarm condition is announced at the monitor. For one or more measurements not configured in the 
Setup VueLink
 menu, an alarm is announced as a text message for the highest priority alarm. 
Priority is determined at the external device.
• always non-latching on the monitor.
• announced as a flashing numeric while the alarm condition persists.
• announced audibly and visibly at the Information Center. 
the monitor is configured to accept external device alarms, 
but the alarms are switched off at the external device.
!
 alarms status of this external measurement is unknown