VENTURE™
PART NO. 1093845
OXYGEN CONSERVED AND BATTERY
USE TIME BASED ON BREATH RATE
i.e. 77% Oxygen conservation over Continuous Flow achieved
with Venture DODD in Variable Breath Mode @ 20 BPM Rate.
Oxygen supply lasts 4.3-times longer. Venture DODD
 
responds to
individual breath rates. Therefore, use times stated are only
approximates. Actual savings depends on individual breathing
patterns.
COMPATIBILITY
Venture DODD requires using a flowrate controlled oxygen
source. Gaseous or liquid oxygen systems that meet the
following requirements are compatible to the Venture
DODD:
Venture 20 requires a 20 +3 psig (138 + 20 kPa) flow source.
Venture 50 requires a 50 +5 psig (345 + 35 kPa) flow source.
Flowrate Range: 0-6 L/Min.
OXYGEN CONSERVATION
BREATH RATE (BPM)
% OXYGEN
 
CONSERVED
10
20
30
40
 Variable Breath Mode:
8277
70
63
 Fixed 1.0 sec. Breath Mode:
83
66
50
33
 Fixed 0.75 sec. Breath Mode:
87
75
6250
 Fixed 0.50 sec. Breath Mode:
91
83
75
66