Emerson 51-DO-03 Benutzerhandbuch
28
MODEL DO-03/04
SECTION 5.0
SOFTWARE CONFIGURATION
5.4 CHOOSING DISPLAY OPTIONS (CONTINUED)
7. Set the remainder of the display parameters. Use the
and keys
to choose the desired parameter. Then press Edit (F4). Use the
key to move the cursor to the desired selection. Press Save (F4) to
store.
store.
SECURITY CAUTION
The analyzer uses the timeout value to activate security. Once
the analyzer is unlocked by entering a security code, security will
not re-activate until a display timeout occurs. If timeout has been
turned off, security will never reactivate.
the analyzer is unlocked by entering a security code, security will
not re-activate until a display timeout occurs. If timeout has been
turned off, security will never reactivate.
Temp units: °C
Output 1: mA
Output 2: mA
Output 1: mA
Output 2: mA
Exit
Edit
Language: English
Display left: I
Display right: Out 2
Display left: I
Display right: Out 2
Exit
Edit
Display contrast: 40
Timeout: On
Timeout value: 10 min
Timeout: On
Timeout value: 10 min
Exit
Edit
Temp units
°C or °F
Output 1
mA or % of full scale
Output 2
mA or % of full scale
Language
English, Français, Español, Deutsch, Italiano
Display left
sensor current (I), alarm 1 setpoint (no units),
alarm 3 setpoint (no units), or blank
alarm 3 setpoint (no units), or blank
Display right
sensor current (I), alarm 2 setpoint (no units),
alarm 3 setpoint (no units), output 2, or blank
alarm 3 setpoint (no units), output 2, or blank
Display Contrast 00 (lightest)-99 (darkest); the display contrast
changes as the number changes
Timeout
Timeout returns the display from any other screen
to the main display if no key is pressed before the
timeout value is exceeded.
to the main display if no key is pressed before the
timeout value is exceeded.
Polling address
Identifies analyzer in multi-drop HART applications.
Timeout value