Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
Section 4: Configuration
3-9000-765 Rev H
May 2015
Using a Field Communicator to configure the meter
 95
5.
Wire Analog Input 1 (AI1)  and Analog Output 1 (AO1) as shown in 
 and 
, drawing DMC-004936. 
Figure 4-20  Field Communicator wiring diagram for the 3810 Series electronics 
  
6.
Use the leads provided with the Field Communicator to connect to your device.
7.
Press and hold the 
Power
 button on the Field Communicator until the green  light 
blinks.
8.
Use the touch screen on the Field Communicator, the keypad or use the stylus to 
navigate through the device menus.
9.
Refer to the Menu tree in Section D.1.1 of the Daniel HART Field Device Specification 
manual (P/N 3-9000-761) for the device fast key sequences. Included in the menu tree 
are:
Diagram Page 1 - 3810 Series Root Menu; 
Overview, Configure|Manual Setup
 
Diagram Page 2 - 
Configure|Manual Setup 
(continued) and 
Alerts Setup
Diagram Page 3 - 
Service Tools|Alerts
 and 
Variables
 
Diagram Page 4 - 
Service Tools|Variables
 (continued), 
Service Tools|Trends
, and 
Service Tools|Maintenance
If you encounter problems, refer to the contact information on the back cover of this manual or 
the contacts included in the Field Communicator Users Manual.