74  Stratos LV/LV-T Technical Manual   
2.3.4 Repetitive Hysteresis 
Repetitive hysteresis is expanded programmability of the 
Hysteresis feature.  Repetitive hysteresis searches for an 
underlying intrinsic cardiac rhythm, which may exist slightly 
below the programmed lower rate (or sensor-indicated rate) of 
the patient.  Following 180 consecutive sensed events, this 
feature allows the intrinsic rhythm to drop to or below the 
hysteresis rate.  During the time when the intrinsic rate is at or 
below the hysteresis rate, pacing occurs at the hysteresis rate 
for the programmed number of beats (up to 10).  Should the 
number of programmed beats be exceeded, the stimulation rate 
returns to the lower rate (or sensor-indicated rate). 
If an intrinsic cardiac rhythm is detected within the programmed 
number of beats between the hysteresis rate and the lower rate, 
the intrinsic rhythm is allowed and inhibits pacing by the CRT-P. 
 
Figure 4.  Repetitive Hysteresis 
Repetitive hysteresis has been incorporated to promote 
spontaneous cardiac rhythm and may reduce device energy 
consumption.  
N
OTES
Repetitive Hysteresis is not active during the programmed 
Night Mode.