11. CALIBRATION 
 
Adjusts the scale for accurate weighing. 
Calibrate the scale in the following cases. 
 
‰
 When the scale is first installed. 
‰
 When the scale has been moved. 
‰
 When the ambient environment has changed.
 
‰
 For regular calibration.
 
 
 
11.1. Gravity acceleration correction   
 
When the scale is first used or has been moved to another location, it should be 
calibrated using a calibration weight. 
But if a calibration weight is not available, the gravity acceleration correction will 
compensate the scale. Change the gravity acceleration value stored in the scale to the 
value of the area where the scale will be used. Refer to the gravity acceleration map at 
the end of this manual. 
 
Note: Gravity acceleration correction is not required when the scale will be 
calibrated using a calibration weight at the place where it is to be used.   
 
‰
 Make sure that the scale is in the weighing 
mode (displaying “kg” (“g”), “lb”, “oz” or “pcs”). 
 
1. Press and hold the  TARE  switch until the 
gravity acceleration value (in this example: 
9.7985) appears, and release the switch. 
 
2. Press the  TARE  switch to enter the 
gravity acceleration value setting mode. 
 
3. Change the displayed value using the following 
switches. 
 UNITS    To shift the blinking digit to the 
right. 
  PRINT      To increase the value of the 
blinking digit by one. 
 
4. Press the  TARE  switch. The setting value 
is stored and “
end
” is displayed.   
5. When calibration using a calibration weight is 
to be performed, go to step 3 of “11.2. 
Calibration using a weight”. To finish the 
setting procedure, turn the power OFF. 
Press and hold
(more than 4 seconds)
Release
Press 
Press
Press